Mark A. White's Selected Abstracts



Mark A. White, Stanley J. Watowich, and Robert O. Fox, (1998), Calibration of Spiral Readout Image Plate Detectors, J. Appl. Cryst. (1999) 32 65-70. {PDF}

An imporved method for intensity uniformity calibration of diffraction data collected on spiral readout image plate (IP) systems is described. The procedure utilizes an attenuated direct beam scan of the IP to generate a radial sensitivity calibration table. Exposure and scanning of the calibration frame is done on the same time scale as typical data collections and requires no additional equipment. Specific examples are presented for use with Mac Science DIP-2000 systems. The improved radial calibration is also shown to lower R merge when the IP is offset from beam center. In addition to improving data quality and statistics this method provides a quick and simple diagnostic tool to monitor changes in the sensitivity of the IP detector as a function of age.